Fluke 726 Precision Multifunction Process Calibrator

FOB Reference Price:Get Latest Price
US $99-9,999 / Pieces | 1 Piece/Pieces (Min. Order)
Report Suspicious Activity
Overview
Quick Details
Place of Origin:
United States
Brand Name:
FLUKE
Model Number:
Fluke 726
Power:
Electronic
Usage:
Process Testing Machine
Voltage DC sources output:
0~1000V, -10mA~75mA
DC Current Source:
0 ~ 24mA
Resistance Sources output:
4000 ohm
Frequency sources output:
15kHZ
SupplyAbility
SupplyAbility:
200 Unit/Units per Month
Packaging & Delivery
Port
shenzhen
Lead Time :
Shipped in 2 days after payment

Fluke 726 Precision Multifunction Process Calibrator :

  • More precise measurement and calibration source performance, accuracies of 0.01%.
  • Two separate channels; measure, source and view process signals simultaneously
  • Measure volts, mA, RTDs, thermocouples, frequency, and resistance to test sensors and transmitters
  • Source/simulate volts, mA, thermocouples, RTDs, frequency, and pressure to calibrate transmitters
  • Measure or source pressure* using any of 29 Fluke 700Pxx Pressure Modules
  • Source mA with simultaneous pressure measurement to conduct valve and I/P tests
  • Integrated pressure switch test allows you to capture the set, reset and deadband of a switch
  • Memory storage for up to 8 calibration results, return stored calibration data from the field for later analysis
  • Transmitter error% calculation, interpret calibration results without a calculator
  • Perform fast linearity tests with auto step and auto ramp features
  • Power transmitters during test using 24 V loop supply and simultaneous mA measurement
  • Store frequently-used test setups for later use
  • Frequency totalizer and frequency pulse train source mode for enhanced flowmeter testing
  • HART mode inserts 250 ohm resistor in mA measure and source for compatibility with HART instrumentation
  • Custom RTD curves, add calibration constants for certified RTD probes for enhanced temperature measurement.
  • New voltage input protection design for improved reliability
  • Three-year warranty

 

------------------------------------------------------------

Measurement Accuracy :

Voltage DC
30.000 V0.01%+ 2 mV (upper display)
20.000 V0.01%+ 2 mV (lower display)
90.00 mV0.01%+ 10μV
-10.00 mV to 75.00 mV0.025 % + 1 count (via TC connector)
Current DC
24.000 mA0.01%+ 2 mA
Resistance
0.0 to 4000.0 Ω+/- 0.015%
Frequency
2.0 to 1000.0 CPM0.05 % + 0.1 cpm
1.0 to 1000.0 Hz0.05 % + 0.1 Hz
1.00 to 15.00 kHz0.05 % + 0.01 kHz
Sensitivity1 V peak-to-peak-minimum
Pressure
Accuracyfrom 0.025% of range using any of 29 pressure modules.
 (for detailed specifications refer to pressure modules in options and accessories)
 Modules available for differential, gauge, vacuum, absolute, dual and high pressure.

 

Source Accuracy :

Voltage DC
100.00 mV0.01%+10 μV
20.000 V0.01%+2 mV
-10.00 mV to 75.00 mV0.025 % + 1 count (via TC connector)
Current DC
24.000 mA (Source)0.01%+ 2 μA
24.000 mA (Simulate)0.01%+ 2 μA
Resistance
Range, Accuracy (Percent of reading + floor) and Excitation Current5 to 400 Ω, 0.015% + 0.1 Ω, 0.1 to 0.5mA
 5 to 400 Ω, 0.015% + 0.05 Ω, 0.5 to 3mA
 400 to 1500 Ω, 0.015% + 0.5 Ω, 0.05 to 0.8 mA
 1500 to 4000 Ω, 0.015% + 0.5 Ω, 0.05 to 0.4 mA
Frequency
2.0 to 1000.0 CPM0.05%
1.0 to 1000.0 Hz0.05%
1.0 to 10.0 kHz0.25%
10.0 to 15.0 kHz0.5%
Waveform:5 V p-p squarewave, -0.1 V offset

 

RTDs and Thermocouples :

Measure and Source Accuracy:
Cu101.8 °C
NI-1200.15 °C
PT-100 (385)0.45 °C
PT-100 (392)0.35 °C
PT-100 (391-JIS)0.35 °C
PT-200 (385)0.95 °C
PT-500 (385)0.55 °C
PT-1000 (385)0.35 °C
RTD Resolution:0.01 C (0.01 F)
Cu10.01 C0.1 C (0.1 F)
J:0.4 °C
K:0.5 °C
T:0.4 °C
E:0.4 °C
R:1.4 °C
S:1.4 °C
B:1.4 °C
L:0.45 °C
U:0.45 °C
N:0.6 °C
XK:0.4 °C
BP:1.1 °C
TC ResolutionJ, K, T, E, L, N, U: 0.1 °C, 0.1 °F B, R, S: 1 °C, 1 °F

 

Specifications :

Ramp Functions
Source functions:Voltage, current, resistance, frequency, temperature
RampsSlow ramp, Fast ramp, 25% step-ramp
Loop Power Function
Voltage:24 V
Accuracy:10%
Maximum current:22 mA, short circuit protected
Step Functions
Source functions:Voltage, current, resistance, frequency, temperature
Steps25% of range, 100% of range

 

Mechanical & General Specifications :

Size
 200mm x 96mm x 47mm (7.9 in x 3.8 in x 1.9 in)
Weight
 650 g (23 oz.)
Batteries
 4 AA alkaline batteries
Battery Replacement
 Separate battery compartment, accessible without breaking calibration seall
Side port connections
 Pressure module connector, also used for remote real-time programming

 

Environmental Specifications :

Operating Temperature
 -10 °C to +55 °C
Storage Temperature
 -20 °C to +71 °C
Operating Altitude
 3000 m

 

Safety Specifications :

Agency Approvals
 EN 61010-1:1993, ANSI/ISA S82.01-1994; CAN/CSA C22.2 No 1010.1:1992

 

 

 

 

 

 

 

 

 

 

 

 

 

 

---------------------------------------------------

Feedback & Shipment: